Data Logger Manual
Page 3
Section 1: Device Safety
CAUTION: DEVICE MUST BE USED IN A MANNER CONSISTENT
WITH THIS MANUAL.
NOTICE: WHEN 230VAC SUPPLY IS USED, DWYER SPECIFIES USE
OF AC SUPPLY PART # T35-9-100R-3 MANUFACTURED BY ENG
ELECTRIC. THIS SUPPLY IS AVAILABLE FROM DWYER.
1.1 Model information
• Model: DLI-T
Description: Measure & record temperature on both an internal and
an external probe channel with LCD display
Manufactured in the USA
1.2 Specifi cations
Internal Temperature
Range: -4 to 140ºF (-20 to +60ºC)
Resolution: 0.2ºF (0.1ºC)
Thermocouple
Types: J, K, T, E, R, S, B, N
Connection: Female subminiature (SMP), fixed screw terminal, or
pluggable screw terminal
Cold Junction Compensation: Automatic, based on internal channel
Thermocouple Type: Range (ºF) Range (ºC) Resolution Accuracy
J
-346 to 1400 -210 to +760 0.1ºC ±0.5ºC
K
-436 to 2498 -260 to +1370 0.1ºC ±0.5ºC
T
-436 to 752 -260 to +400 0.1ºC ±0.5ºC
E
-436 to 1796 -260 to +980 0.C ±0.5ºC
R
-58 to 3200 -50 to +1760 0.5ºC ±2.0º C
S
-58 to 3200 -50 to +1760 0.C ±2.0º C
B
140 to 3308 +60 to +1820 0.5ºC ±2.0º C
N
-436 to 2372 -260 to +1300 0.1ºC ±0.5ºC