Series UT TransmittersSpecifications - Installation and Operating InstructionsBulletin L-UTDWYER INSTRUMENTS, INC. Phone: 219/879-8000 www.dwyer-inst.
Setting the Series UTC Relay ValuesYou can set the relays to five different configuration modes: Level, Distance andFlow (where applicable), Volume an
Setting Relay 4 to Report Errors This mode enables you to use the relay as a trigger to set on an alarm or siren incase the unit produces inaccurate m
To Set the Relay Values for Level or Distance:Setting the 20 mA/4 mA LevelsSeries UT enables you to set height, volume or flow values to be used as 20
Setting the Application TypeThe Series UT main menu displays (by default) LIQUID for models intended forliquid. Series UT model intended
The scan distance function is accessed from the default screen. AUTO. CAL isdisplayed at the base of the display screen during the
To Select the Flow Measurement Settings:Open Channels Flow MeasurementsThe flume/weir type code methodology used when setting up open channe
Trapezoidal (Cipolletti) Sharp-Crested Weir (Type 3)Khafagi-Venturi Flume (Type 5)Parshall Flume (Type 6)Palmer Bowlus Flume Trapezoidal Throat Cross-
H Flume (Type 8)Neyrpic Venturi Flume/Long-Base Weir (Type 9)Long-Base WeirFlumes/Weirs - American StandardRectangular Suppressed Sharp-Crested Weir (
V-Notch (Triangular) Sharp-Crested Weir (Type 4)Parshall Flume (Type 5)Palmer Bowlus Flume Trapezoidal Throat Cross-Selection (Type 6)H Flume (Type 7)
Leopold-Lagco Flume(Type 8)Chapter 5: Additional FeaturesThis chapter describes the functions available in Series UT’s additional menu.The additional
Table of ContentsChapter 1. Introducing Series UTSensor Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Appendix A: Series UT RangesThis appendix provides the maximum and minimum permitted values for the mainmenu functions in the Series UTS and Series UT
Chapter 1: IntroductionSeries UT is a non contact, ultrasonic, continuous level measurement instrumentthat is able to provide accurate meas
SPECIFICATIONSSERIES UTCService: Compatible liquids.Wetted Materials: ECTFE.Ranges: (Sensor Dependent)Short Range: 1.3 to 39.4 ft (0.4 to 12 m);Standa
Flexible CableFlexible sensor cables are supplied with 16.4 ft/5 m loose-end cable. However, theycan be modified to fit longer length than 16.4 ft/5 m
Sensor PositioningWhen installing the sensor, ensure that it is:• Mounted above the dead-zone area (Blind Zone) (blocking distance). • Positioned at l
Wiring the Sensor CableSensor cable specifications: 5 wires, 0.75 mm2overall shielded.After the sensor is installed on the tank, the free en
Using the Series UT Function ButtonsThe function buttons are used to perform various operations, summarized in thefollowing table. Modifying N
temporarily while Series UT is taking a reading. You can refresh the reading by pressing the ENT and BACK buttonssimultaneously. Setting Main